Sample: entirely fictitious example, not a client case.

Based on GAMP 5 Second Edition and EU GMP Annex 11. Professional assessment, not legal advice.

1 Framework

Company
Mid-sized testing laboratory that analyses samples for pharmaceutical manufacturers and works under GMP.
Occasion
A new laboratory information management system (LIMS) is due to go live on [date per project plan].
Markets
EU, so Annex 11 applies. Whether customers also use the results for the USA is open.
Existing
A procedure for computer system validation, but no validation master plan.
Criticality
High with direct impact on patient safety or product quality, medium with impact on data integrity or indirect impact, otherwise low.

2 Classification of the systems

SystemCriticalityClassification
LIMShighGAMP category 4, configured software. The test results feed into the customers' release decisions. The supplier has been assessed by questionnaire and documents, not audited.
Interface from the chromatography data system to the LIMSmediumCategory 5, developed specifically for the laboratory. It transfers results; an error corrupts data.
LIMS server and databasemediumCategory 1, infrastructure. It is qualified, not validated.
Invoicing in the ERPnot GxPNo impact on product quality or test data.

3 Risk-based validation scope

LIMS

  • Validation plan, user requirements and functional risk assessment.
  • Configuration specification and traceability from requirements to tests.
  • Supplier audit before go-live, because criticality is high.
  • Audit trail, access control and backup as requirements of their own.

Interface

  • Requirements plus functional and design specification.
  • Code review and unit tests at the developer, plus a supplier audit.

Server and database

  • Document installation and configuration, that is, qualify.
  • Operation under change control.

Invoicing

  • No GMP validation. The classification is justified and reviewed if the use changes.

4 Plan for tests, evidence and approval

TestForDepth
Installation qualification (IQ)All three GxP systemsIn the target environment.
Operational qualification (OQ)LIMSAll GxP-critical functions with positive and negative tests, independently reviewed. Supplier tests are credited where assessed.
Operational qualification (OQ)InterfaceCritical functions per the risk assessment.
Audit trailroles and rightsLIMSIncluding backup and restore.
Performance qualification (PQ)LIMS with interfaceWith real workflows and the future users.
Report and approvalAll three GxP systemsClose deviations, approval by the quality unit, handover to operations.

Deliberately not: invoicing is not validated, and qualification is enough for server and database. That keeps the effort where test data is created.

5 Open points

  • Confirm [date per project plan] with the supplier and the quality unit.
  • Clarify whether customers use the results for the US market. If so, 21 CFR Part 11 also applies to electronic signatures.
  • Whether supplier tests can be credited depends on the assessment of the supplier's documents.

6 Next steps

  1. Draw up the validation plan and have the quality unit approve it.
  2. Write requirements and risk assessment for LIMS and interface.
  3. Carry out supplier audits at the LIMS supplier and the interface developer.
  4. Run the tests according to plan, assess and close deviations.
  5. After approval, set the periodic review.

Reservation

This sample is a professional assessment and not legal advice. It is based on GAMP 5 Second Edition and EU GMP Annex 11, and for the US market 21 CFR Part 11. Scope and test depth are a proposal that the quality unit sets and approves in the validation plan. Values in square brackets appear in the real report as results of the assessment.